Attach the current input of one of the lock-ins to another contact as the drain

Trigger the second transport lock-in off of the first one. Attach the output of one of the lock-ins to the 1/10 voltage divider, then to a contact of the sample. You can attach the voltage contacts somewhere if you want to, this is not particularly important though. It may be necessary to a apply a voltage to the gates, especially if you are working with semiconducting materials, like the transition metal dichalcogenides.Increase the voltage until you see 1 µA of current. In my experience, members of the nanoSQUID team tend to be a little too timid about applying large currents to these samples because they are very susceptible to damage through electrostatic discharge, and of course it feels pretty bad to damage a device somebody else made for you. Although it’s true that researchers doing transport measurements almost never use currents as high as 1 µA, I can tell you that we have never damaged a heterostructure with high current at all, large pot with drainage and certainly not at 1 µA. It is generally pretty safe to go as high as 100 µA, and we have gone considerably higher than that too.

Currents greater than 1 µA will saturate the lock-in input, but you can still increase the voltage if you need to . Alternatively you can use the Ithacos adjustable transimpedance amplifier as the sink. If you do so, be careful not to adjust any of the knobs on this device while it is hooked up to the heterostructure, because adjusting those knobs can produce pulses of current large enough to damage devices. While you’re increasing the voltage, keep an eye on the SQUID signal. Increase the time constant if it helps you see the signal . Once you see a signal on the nanoSQUID channel of the Zurich, set up a scan. Check that the auxiliary outputs from the Zurich are going to the right ADC inputs, the right ADC inputs are correctly labelled in the scan window, the right auxiliary outputs are set up in the Zurich ‘Aux’ window and are sampling the right channels, and the right channels are set up and activated in the Zurich window. You should definitely see a thermal gradient if the signal is 3-5x the noise floor. If you don’t, I’d recommend investing some time into making sure the measurement is set up correctlyyou don’t want to just keep increasing current through the sample in response to not seeing features on a scan that isn’t set up right! If you get really frustrated and want a sanity check, click “Set Position” to each of the corners of the scan range and watch the signal on the Zurich control panelit should change if everything is working. There are a lot of issues that can affect scanning, and it isn’t really possible to cover all of them in this document, so you will have to rely on accumulated experience.

Some problems will become obvious if you just sit and think about them- for example, if the thermal gradient is precisely along the x-axis and coarse positioner navigation is failing to find a strong local maximum it likely means that the y-axis scanner is disconnected or damaged. In Andrea’s lab, the basic circuits on the 1.5K and 300 mK systems as currently set up should be pretty close to working, so if there’s a problem I’d recommend observing the relevant circuits and thinking about the situation for at least a few minutes before making big changes. The scanners as currently installed on the 1.5K system do not constitute a healthy right-handed coordinate system, so to navigate you will need a lookup table translating scanner axes into coarse positioner axes. I think this issue is resolved on the 300 mK system, but this is the kind of thing that can get scrambled by upgrades and repair campaigns. In all of our note taking Powerpoints and EndNotes, we have a little blue matrix that relates the scan axes to the coarse positioner axes. Use this to determine and write down the direction you need to move in the coarse positioner axes in your notes. You now have an initial direction in which you can start travelling. We will next perform long distance thermal navigation, at a height of 150 µm above the surface. Retract 150 µm using axis 3 of the coarse positioners. I’d recommend doing this in one or two big steps, because the coarse positioner can slide in response to small excursions. Verify that you can still see the thermal signal on the SQUID. It is Ok if it’s faint or close to the noise floor; it will increase in size, and you know which directions to start travelling. If the resistive encoders are working , then use them to move in 100 um steps, checking the SQUID signal in between movements. There is no need to ground the SQUID in between coarse positioner steps, there will be crosstalk but this is not hazardous for the nanoSQUID. If the resistive encoders are not working, click the Step+ button repeatedly until the SQUID signal increases to a maximum.

This might take a few minutes or so of clicking. You can work on a software solution instead if you like , but remember that there is always a simple, safe solution available! Once the signal is at a maximum, take another scan to verify that you’re centered above the device. You should see a local maximum in the temperature in the middle of your scan region. Ground the SQUID. Ramp the current through the device down to zero. Zero and ground any gates you have applied voltages to. Ground the sample. Make sure the SQUID is grounded to the breakout box by a BNC . Hook up the second little red turbo pump to the sample chamber through a plastic clamp and o-ring, and turn it on. Slowly, over 10-20 minutes, drainage collection pot open the valve to the sample chamber and pump it out. Make sure the sand buckets for vibration isolation are set up and the bellows aren’t touching the ground. If there are vibration issues you can often feel them on the bellows and on the table with your hand. Repeat the setup for approaching to contact, and approach to contact. Definitely watch the first few rounds of this approach! You can even watch the whole thing- it’ll take 30-45 mintues, but if you’ve messed something up then the approach will destroy both the SQUID and the device, because you’ve carefully aligned the SQUID with the device! Once you’ve reached the surface, you will set up the SQUID circuit. Attach the preamplifier to one of the SMA connectors at the top of the insert. Attach its output to the input of the feedback box. This output goes through the ground breaker that is clamped to the table in Andrea’s lab; all of these analog electronic circuits are susceptible to noise and ringing, so I’m sure there will be different idiosyncracies in other laboratories with other electromagnetic environments. Attach the output of the feedback box to the BNC labelled FEEDBACK . This is the BNC that should get a resistor in series if you wanted to increase the transfer function. We generally use resistors between 1 kΩ and 10 kΩ for this. To start with, just using nothing is fine . Plug the preamp and feedback box into fresh batteries . Turn the preamp on. Turn the feedback OFF. Hook up the SQUID bias wires to SQUID A and SQUID B. You can tell which they are because of the chunky low pass filters on the end, but of course they are also labelled. Make sure both sides of the SQUID are grounded while hooking it up- there is a BNC T there for a grounding cap for this purpose. Hook up Output 2 of the Zurich to signal input on the feedback box. Apply 1 V to signal input. There’s a good chance you just used this same output and cable to apply avoltage to the device, so be careful not to skip this step and apply this voltage to the device itself! You should see the SQUID array transfer function on the oscilloscope . Turn the rheostat/potentiometer on the preamp until this pattern has maximum amplitude. Turn the Offset rheostat/potentiometer on the feedback box until this passes through zero . There is a more sophisticated procedure for minimizing noise in the SQUID array; this is covered in great detail by documents Martin Huber has provided to the lab. But if you are a beginner this simple procedure will work fine. Flip the On switch on the feedback box, and watch the interference pattern vanish, replaced by a line near V = 0. Turn off the AC voltage going to signal input. You are now ready to characterize the SQUID, although you’ll need to unground it. That includes removing the BNC grounding caps from the T’s downstream of the SQUID bias filters and also flipping the BNC switch on the top of the rack. Click ‘preliminary sweep’ on the nSOT characterizer window. Sweep from 0 to 0.1. If you see a linear slope, a ton of stuff is working!

The SQUID bias circuit, the SQUID array, the feedback electronics, all the cryogenics- that’s a really good sign. If you see no signal, don’t panic. Once again, there’s a lot of stuff involved in this circuit and a ton of mistakes you can make. Go back through the list and check everything, then check to make sure the SQUID bias isn’t grounded somewhere. Increase the sweep range until you see a critical current or you get above 3.3 V, which is where the feedback box will fail. If you don’t see a critical current, you have a SHOVET but not a SQUID. If you see a critical current, close the window, switch to the nSOT characterizer, and characterize the SQUID. At this point, you are at the surface and over the device with a working SQUID, and you can begin your imaging campaign, so what comes next is up to you!Anthocyanins constitute a large family of plant polyphenols and are responsible for many of the fruit and floral colors observed in nature. Anthocyanins are water-soluble pigments located in the grape skin vacuoles that, during the fermentation process, are released into the wine. It has been demonstrated that determining the amount of pigments present in the berries is not enough to estimate the concentration of anthocyanin in the final product. This lack of correlation is mainly attributed to the interaction between the pigments and the skin cell walls during the extraction process. Additionally, the adsorption of phenolics to solids in the fermentor after being released, such as grape skins and yeast hulls, has previously been demonstrated. Previous studies have shown that the interaction between polyphenols and skin cell walls is dependent on the composition of the latter suggesting that specific cell wall constituents show different adsorption capacities for polyphenols. In the case of anthocyanins, the cellulose content and the degree of methylation of the pectin have shown positive correlations with the adsorption capacity. Moreover, some studies suggest that other components of the cell walls, such as proteins, can occupy binding sites resulting in overall lower anthocyanin adsorption.Another factor that greatly influences the extraction of phenolics during wine fermentation has been shown to be the temperature at which the fermentation is performed. Previous research has shown that elevated fermentation temperatures produce finished wines that are more highly colored and have greater concentrations of pigmented polymers. The increase in extracted phenolics at elevated temperatures has been accredited to two temperature related effects: an increased permeability of the hypodermal cells of the grape skins and an increase in the solubility of phenolics at higher temperature. It has also been shown that changes on the temperature can impact the physical structure of the cell wall material . It has been postulated that, at high temperatures, the cellulose structure opens up, potentially creating new sites and a faster exchange between the molecules. Additionally, an increase in temperature can disrupt hydrogen bonds between the cell wall and the phenolics increasing its concentration in solution. A second fermentation factor that is also likely to have a significant effect is the production of ethanol during fermentation.